You are required to read and agree to the below before accessing a full-text version of an article in the IDE article repository.

The full-text document you are about to access is subject to national and international copyright laws. In most cases (but not necessarily all) the consequence is that personal use is allowed given that the copyright owner is duly acknowledged and respected. All other use (typically) require an explicit permission (often in writing) by the copyright owner.

For the reports in this repository we specifically note that

  • the use of articles under IEEE copyright is governed by the IEEE copyright policy (available at http://www.ieee.org/web/publications/rights/copyrightpolicy.html)
  • the use of articles under ACM copyright is governed by the ACM copyright policy (available at http://www.acm.org/pubs/copyright_policy/)
  • technical reports and other articles issued by M‰lardalen University is free for personal use. For other use, the explicit consent of the authors is required
  • in other cases, please contact the copyright owner for detailed information

By accepting I agree to acknowledge and respect the rights of the copyright owner of the document I am about to access.

If you are in doubt, feel free to contact webmaster@ide.mdh.se

Impact of electrode misplacement on the accuracy of 12-lead ECG synthesized from differential leads

Authors:

Ivan Tomasic, Roman Trobec, Aleksandra Rashkovska , Maria Lindén

Publication Type:

Conference/Workshop Paper

Venue:

The 38th international convention on information and communication technology, electronics and microelectronics

Publisher:

IEEE

DOI:

10.1109/MIPRO.2015.7160296


Abstract

The sensitivity to electrode misplacement of the 12-lead ECG, synthesized from differential leads, was analyzed by simulating the electrode misplacement, with a help from interpolated body surface potential maps. The electrodes forming optimal differential leads were independently moved only on the lines defined by gradient vectors, calculated on the integral body surface potential map, obtained from juxtaposed average beats for each of the 47 multichannel ECGs used in the study. The 12-lead ECGs were synthesizes from the simulated moved differential leads and compared to the ECGs synthesized from the nominal electrodes’ positions, by using correlation coefficients and RMSD. The results show that synthesized 12-lead ECGs are less accurate if electrodes with bigger associated gradients are misplaced, but also suggest that small displacements of up to 1 cm do not have a big influence on the synthesis quality. From the perspective of ECG synthesis, the presented methodology can be used to define an additional criterion for optimal triplets of differential leads selection, which will ensure that the differential lead triplet are optimal not only in terms of the quality of the synthesized leads, but also in terms of being minimally sensitive to electrode misplacement.

Bibtex

@inproceedings{Tomasic3916,
author = {Ivan Tomasic and Roman Trobec and Aleksandra Rashkovska and Maria Lind{\'e}n},
title = {Impact of electrode misplacement on the accuracy of 12-lead ECG synthesized from differential leads},
pages = {365--369},
month = {June},
year = {2015},
booktitle = {The 38th international convention on information and communication technology, electronics and microelectronics},
publisher = {IEEE},
url = {http://www.es.mdu.se/publications/3916-}
}